TI-2000 Medical Hospital Equipment Transport Baby Infant Incubator
Standard configuration:
Main body (including the Acrylic glass hood, temperature controller, bassinet, internal battery, and observation lamp), oxygen cylinders, oxygen supply system, skin temperature sensor, I. V. Pole, mattress, adjustable stand)
Optional configuration: >37° C override temperature mode
Specification
N AC Power supply: AC220V-230V, 50Hz
N DC Power supply: DC12V/10A or DC24V/6A
N Input power: 400VA
N Mode of control: Air mode and baby mode controlled by micro-computer
N Control range of air temperature: 25° C- 37° C(Override mode 37-38° C)
N Control range of baby temperature: 34° C- 37° C(Override mode 37-37.5° C)
N Accuracy of skin temperature sensor: ≤ 0.3° C
N Variability of temperature: ≤ 1.0° C
N Uniformity of temperature: ≤ 1.5° C
N Noise inside hood: ≤ 60dB(A)
N Failure alarm: Over-temperature alarm, temperature deviation alarm, sensor alarm, Motor Fan Failed Alarm, Power failure Alarm and so on.
N Running time of internal battery: 90min (1 internal battery)AC Power supply AC220V-230V, 50Hz DC Power supply DC12V/10A or DC24V/6A Input power 400VA Control range of air temperature 25° C- 37° C(Override mode 37-38° C) Accuracy of skin temperature sensor ≤ 0.3° C -