HB6601 glycated hemoglobin hba1c analyzer
Inside Heating
Durable Pump Tubing
Filter for avoiding blockage
Pressure Monitor
Basic Parameters
Testing Method
Low pressure ion exchange liquid-phase chromatography
Testing Item
Glycosylated hemoglobin A1c
Testing Scope
3% to 18%
Testing Parameters
Precision (CV) ≤3%, Accuracy (V) ≤1.50%
Testing Time
4 minutes (reporting), Full-processing 6 minutes (including cleaning column and recovery time)
Sample Type
Venous blood (EDTA anticoagulation), Peripheral finger blood
Sample volume
5μl full-blood for each of samples
Functional Parameters
Photometer
415nm LED Integral flow colorimeter
Sampling mode
Full-auto 25-position sample turntable (20 sample positions, 2 quality control positions, 1 emergency position, 1 cleaning position, 1 zero position)
Testing mode
Batch auto test, single test for emergency
Calibration
Optional selection of 2 points or 4 points, optional selection of manual or automation Equipped with 2 groups of high and low level calibrators
Thermostatic Control
Column and agent thermostatic control: 25 ± 0.5°C
Reagent Matching
One piece of column supplied with consumption/300 pieces of testing samples
Reagent Allocation
Each of standard agent matched with A, B, C, D eluent, calibrator, hemolytic agent, pump pipe
Output Parameters
Display
320x240 LCD Graphic Display, displaying real-time testing cure
Printer
58mm Thermal printer, printing testing cure and report
Report Output
IFCC concentration value, NGSP area percentage, ADAG average glucose
Data storage
1000 test report (including testing cure)
Communication interface
RS232 communication interface, connecting to HISILIS system
Working Parameters
Power Supply
AC220 V ± 22V 50Hz 65VA
Overall size
408mm × 390mm × 230mm
Weight
8Kg
Working Environment
Temperature 12°C-30°C,Relative humidity ≤85% -
GLYCATED HEMOGLOBIN HBA1C ANALYZER
Model NO.: HB6601Type: Biochemistry AnalyzerDisplay: 320X240 LCD Graphic Display, Displaying Real-TiSample Volume: 5UL Full BloodTesting Mode: Batch Auto Test, Single Test for EmergencySpecification: ISOClassification: Biochemical ApparatusCertification: ISO13485Sample Type: Venous Blood (EDTA Anticoagulation), PeripheralSampling Mode: 25 Position Sample Turntable PositionOrigin: Jiangsu, China